• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI CAPSURE SENSE LEAD MRI SURESCAN; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI CAPSURE SENSE LEAD MRI SURESCAN; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 407458
Device Problems Break (1069); High impedance (1291); Low impedance (2285); Device Contamination with Body Fluid (2317); High Capture Threshold (3266)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/17/2022
Event Type  Injury  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during attempted implant the right atrial (ra) lead exhibited high thresholds in multiple locations.The ra lead was attempted/not used.It was also reported that approximately one month later during attempted implant of a new ra lead the chronic right ventricular (rv) lead was retested and it exhibited very high thresholds.It was also reported that the rv lead exhibited low and high fluctuating impedance.It was also reported that on inspection of the rv lead it was noted that the insulation was damaged in two locations that allowed significant blood ingress into the insulation layers.The rv lead was explanted and replaced. no patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the full lead was returned and analyzed.Analysis indicated the inner insulation of the lead was extrinsically breached due to a tear.The outer insulation of the lead was extrinsically breached due to a cut.The proximal conductor of the lead became extrinsically distorted due to kinking/buckling.The distal conductor was extrinsically distorted due to kinking/buckling.Visual analysis of the lead indicated apparent explant damage.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSURE SENSE LEAD MRI SURESCAN
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key15671990
MDR Text Key302397981
Report Number2649622-2022-21543
Device Sequence Number1
Product Code DTB
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P830061
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/26/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number407458
Device Catalogue Number407458
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/16/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/15/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
W3DR01 IPG, 6725 ADAPTOR
Patient Outcome(s) Hospitalization; Required Intervention;
-
-