Model Number M00535900 |
Device Problem
Device Contaminated During Manufacture or Shipping (2969)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/01/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: approximated based on the date the manufacturer became aware of the event.(b)(4).
|
|
Event Description
|
It was reported to boston scientific corporation that an ultratome xl was viewed at the "time of the conference in the consigned sector".The exact event date was unknown.Upon checking the device, a hair was found inside the package.There was no procedure involved and the device was not used in the patient.No further information has been obtained despite good faith efforts.
|
|
Manufacturer Narrative
|
Block b3: approximated based on the date the manufacturer became aware of the event.Block h6 (device codes): medical device problem code a1802 captures the reportable event of foreign material present in the packaging of the device.Block h10: the returned ultratome xl was analyzed, and a visual evaluation noted that a foreign material was inside the pouch, which was consistent with the findings when the pouch was observed under magnification.No other problems with the device were noted.The reported event of foreign material inside the package was confirmed.Upon analysis, it was found that a foreign material was found inside the pouch.Based on all gathered information, this complaint will be document as manufacturing deficiency.An investigation is in place to address this problem.
|
|
Event Description
|
It was reported to boston scientific corporation that an ultratome xl was viewed at the "time of the conference in the consigned sector".The exact event date was unknown.Upon checking the device, a hair was found inside the package.There was no procedure involved and the device was not used in the patient.No further information has been obtained despite good faith efforts.
|
|