Model Number M00535900
Device Problem Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/01/2022
Event Type  malfunction  
Manufacturer Narrative
Date of event: approximated based on the date the manufacturer became aware of the event.(b)(4).
 
Event Description
It was reported to boston scientific corporation that an ultratome xl was viewed at the "time of the conference in the consigned sector".The exact event date was unknown.Upon checking the device, a hair was found inside the package.There was no procedure involved and the device was not used in the patient.No further information has been obtained despite good faith efforts.
 
Manufacturer Narrative
Block b3: approximated based on the date the manufacturer became aware of the event.Block h6 (device codes): medical device problem code a1802 captures the reportable event of foreign material present in the packaging of the device.Block h10: the returned ultratome xl was analyzed, and a visual evaluation noted that a foreign material was inside the pouch, which was consistent with the findings when the pouch was observed under magnification.No other problems with the device were noted.The reported event of foreign material inside the package was confirmed.Upon analysis, it was found that a foreign material was found inside the pouch.Based on all gathered information, this complaint will be document as manufacturing deficiency.An investigation is in place to address this problem.
 
Event Description
It was reported to boston scientific corporation that an ultratome xl was viewed at the "time of the conference in the consigned sector".The exact event date was unknown.Upon checking the device, a hair was found inside the package.There was no procedure involved and the device was not used in the patient.No further information has been obtained despite good faith efforts.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRATOME XL
Type of Device
SNARE, FLEXIBLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15672741
MDR Text Key307095750
Report Number3005099803-2022-06126
Device Sequence Number1
Product Code FDI
UDI-Device Identifier08714729103257
UDI-Public08714729103257
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K930022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/26/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00535900
Device Catalogue Number3590
Device Lot Number0029411723
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received01/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/16/2022
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-