Model Number 150075
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Chest Pain (1776); Low Blood Pressure/ Hypotension (1914); Pain (1994); Cardiac Tamponade (2226); Cardiac Perforation (2513); Insufficient Information (4580)
Event Date 09/19/2022
Event Type  Injury  
Event Description
An attempt was made to use one 7f pulmonary wedge pressure (pwp) catheter.It was reported that following successful implantation of a non-medtronic pulmonary artery pressure monitoring system it was found that the right ventricle had been perforated and the patient had a pericardial effusion.The patient began complaining of back pain and chest pain and the blood pressure began to drop.A computed tomography (ct) scan and echo were performed and the pericardial effusion was identified.A pericardiocentesis procedure was performed to drain the effusion.The patient was brought to the intensive care unit for monitoring, but remained unstable.The patient was taken for open heart surgery to assess/repair the source of the pericardial effusion.The perforation injury in the right ventricle was found and successfully repaired.It was detailed that it could not be confirmed if the perforation was caused by the medtronic pwp catheter or by the guidewire used in the initial procedure.It was believed that it was not caused by the non-medtronic pulmonary artery pressure monitoring system due to the location of the perforation in the right ventricle.No further patient injury was reported and the patient is recovering from the surgery.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Annex d code.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Manufacturer Narrative
Additional information: there were multiple wires and 2 catheters used during the procedure.Patient past medical history provided.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PWP CATHETER
Type of Device
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
MEDTRONIC, INC
37a cherry hill dr
danvers MA 01923
Manufacturer (Section G)
MEDTRONIC, INC
37a cherry hill dr
danvers MA 01923
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key15672764
MDR Text Key302421535
Report Number1220452-2022-00079
Device Sequence Number1
Product Code DQO
UDI-Device Identifier00613994733658
UDI-Public00613994733658
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K884065
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 12/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/26/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number150075
Device Catalogue Number150075
Device Lot Number0011311390
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/15/2022
Date Device Manufactured07/13/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-