• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP CRYOCONSOLE; CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP CRYOCONSOLE; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Model Number 106A3
Device Problems Infusion or Flow Problem (2964); Protective Measures Problem (3015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/21/2022
Event Type  Injury  
Event Description
It was reported that during a cryo ablation procedure, a system notice was received indicating that the refrigerant delivery path was obstructed.Additionally, the low pressure regulator (lpr) and high pressure regulator (hpr) values were not as expected.The coaxial umbilical cable and balloon catheter were replaced without resolution.Also, the console was vented without resolution. the case was aborted under general anesthesia. service was performed at a later date.The low pressure regulator (lpr) and high pressure regulator (hpr) were adjusted which resolved the issue.The no patient complications have been reported as a result of this event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Product event summary: data files were returned and analyzed.Patient files showed at least eight applications were performed with console 106a3 with serial number n6862 on the date of the event, and confirmed system notice 50012 (indicating the refrigerant delivery path is obstructed) in injections one, two, and eight on the date of event.In conclusion, the reported issue (aborted under general anesthesia) cannot be assessed through data analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRYOCONSOLE
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key15677128
MDR Text Key302468900
Report Number3002648230-2022-00412
Device Sequence Number1
Product Code LPB
UDI-Device Identifier00643169195844
UDI-Public00643169195844
Combination Product (y/n)N
PMA/PMN Number
P020045
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/26/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number106A3
Device Catalogue Number106A3
Was Device Available for Evaluation? No
Date Manufacturer Received01/18/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-