MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number HL 20
Device Problem Pumping Problem (3016)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/05/2022
Event Type  malfunction  
Manufacturer Narrative
During service belt slip error message on the pump module of hl 20 was reported.No harm to any person has been reported.A getinge field service technician was onsite and investigated the unit in question.It was found that the optical tacho board was defective.The problem was resolved by replacing the optical tacho board.The system was checked according to factory¿s specification and all tests passed.The device was put back into use.Thus the reported failure could be confirmed.The defective part has been scrapped by the customer, therefore no technical investigation could be performed.However a defective optical tacho board is a plausible cause for the error message "belt slip".Because the sensor no longer detects one or more lines on the film, it measures an apparently low speed at the pump head.Since this speed is then lower than the motor speed, the pump interprets this as a slipping belt and displays this message.The failure mode "belt slip" error message can also be linked to the following most possible root causes according to the hl 20 risk management file: -defective/ dirty tacho, relay or pump belt device was manufactured in 2013-10-29.The review of the non-conformities during the period of 2013-10-29 to 2022-10-20 does not show any non-conformity in regard to the reported product and failure.There is no indication that manufacturing issues occurred during this time, thus production related influences are unlikely.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
During service belt slip error message on the pump module of hl 20 was reported.No harm to any person has been reported.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART LUNG MACHINE
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key15678548
MDR Text Key304345498
Report Number8010762-2022-00423
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K943803
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Consumer,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/27/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL 20
Device Catalogue Number701043262
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/18/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/29/2013
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-