Model Number PCF-H190DL |
Device Problem
Device Handling Problem (3265)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/05/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The air/water (a/w) channel cleaning adapters were ordered by the facility on urgent basis.In the meanwhile the olympus endoscopy support specialist (ess) advised the operating room staff to use the a/w channel cleaning adapters form the gastrointestinal department.The device is not returned, as such a definitive root cause of the reported complaint cannot be determined at this time.This event is under investigation.A supplemental report will be submitted upon completion of the investigation or upon receiving additional information.
|
|
Event Description
|
An olympus endoscopy support specialist (ess) performed a repair reduction in-service for the operating room staff by observing procedures and reprocessing for repair trends and ways to avoid and prevent common repairs.During pre-cleaning, ess observed that the air/water (a/w) channel cleaning adapter was used.The facility did not have the a/w channel cleaning adapter, nor was a disposable a/w channel cleaning adapter being used.Ess trained the staff for pre-cleaning steps per the instructions for use (ifu) including the importance of using an a/w channel cleaning adapter.There is no reported harm to any patient.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause could not be established.The suggested phenomenon was presumed to have been due to the user¿s understanding differing from the olympus recommendation in device handling and reprocessing steps.The instructions for use (ifu) (reprocessing manual) warns against insufficient reprocessing as follows: "precautions: an insufficiently cleaned, disinfected or sterilized endoscope and/or accessories may pose an infection control risk to the patients and/or operators who contact them." the ifu (reprocessing manual) states the method of flushing the air/water channel with water and air and the method of flushing the auxiliary water channel at ¿5.3 precleaning the endoscope and accessories¿.Olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|