Model Number PCF-H190DL
Device Problem Device Handling Problem (3265)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/05/2022
Event Type  malfunction  
Manufacturer Narrative
The air/water (a/w) channel cleaning adapters were ordered by the facility on urgent basis.In the meanwhile the olympus endoscopy support specialist (ess) advised the operating room staff to use the a/w channel cleaning adapters form the gastrointestinal department.The device is not returned, as such a definitive root cause of the reported complaint cannot be determined at this time.This event is under investigation.A supplemental report will be submitted upon completion of the investigation or upon receiving additional information.
 
Event Description
An olympus endoscopy support specialist (ess) performed a repair reduction in-service for the operating room staff by observing procedures and reprocessing for repair trends and ways to avoid and prevent common repairs.During pre-cleaning, ess observed that the air/water (a/w) channel cleaning adapter was used.The facility did not have the a/w channel cleaning adapter, nor was a disposable a/w channel cleaning adapter being used.Ess trained the staff for pre-cleaning steps per the instructions for use (ifu) including the importance of using an a/w channel cleaning adapter.There is no reported harm to any patient.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause could not be established.The suggested phenomenon was presumed to have been due to the user¿s understanding differing from the olympus recommendation in device handling and reprocessing steps.The instructions for use (ifu) (reprocessing manual) warns against insufficient reprocessing as follows: "precautions: an insufficiently cleaned, disinfected or sterilized endoscope and/or accessories may pose an infection control risk to the patients and/or operators who contact them." the ifu (reprocessing manual) states the method of flushing the air/water channel with water and air and the method of flushing the auxiliary water channel at ¿5.3 precleaning the endoscope and accessories¿.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III COLONOVIDEOSCOPE
Type of Device
COLONOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer Contact
masaharu hirose
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8-520
JA   965-8520
426422891
MDR Report Key15685279
MDR Text Key306800850
Report Number9610595-2022-03405
Device Sequence Number1
Product Code FDF
UDI-Device Identifier04953170363672
UDI-Public04953170363672
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131780
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/27/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPCF-H190DL
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/28/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/19/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-