The mayfield skull clamp (a1059) was returned for evaluation: device history record (dhr) - the dhr was reviewed, and no anomalies related to the reported failure was observed.Failure analysis - the investigation of the returned device found no device deficiencies that would have contributed to the reported complaint.Service & repair (s&r) could not duplicate the springing back of the pressure screw or ratchet arm as described.The serration of the arm was in good condition; however, to be on the safe side, the plunger assembly that fixes the ratchet arm was replaced as a preventive measure.There was some play in the lock assembly along with worn components that were replaced.The skull clamp was then checked for functionality.Root cause -probable root cause of the reported incident is improper or suboptimal placement of the skull clamp.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
|