|
Model Number 1036 |
Device Problems
Low impedance (2285); Material Twisted/Bent (2981)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/22/2022 |
Event Type
Injury
|
Event Description
|
A patient had a upper back and neck chiropractic adjustment in (b)(6) of 2022.On (b)(6) 2022, records showed a drop in impedance occurred on (b)(6) 2022 which correlated with the adjustment.X-rays were performed, and the csl lead insulation appeared torn at the mid-neck level.The csl was also twisted and wrapped around itself, which indicated that the patient twiddled with their device.It was noted that the ipg had been sutured with two sutures at the initial implant procedure.A csl replacement occurred on (b)(6) 2022, and it was noted that the ipg sutures were still intact on the ipg but had pulled out of the tissue.A new ipg was placed subpectorally and resutured.The patient was doing well following the procedure.
|
|
Manufacturer Narrative
|
The ipg and a portion of the csl were returned for analysis.The ipg functioned as intended.No analysis was able to be performed on the csl, as only a portion was returned, so csl analysis was inconclusive.Based on the information received, the cause of the reported event was related to twiddling and a chiropractic adjustment.The device history record and sterilization record for these device serial numbers have been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The devices met material, assembly, and quality control requirements.Cvrx id# (b)(4).
|
|
Search Alerts/Recalls
|
|
|