ROCHE DIAGNOSTICS ACCU-CHEK INFORM II TEST STRIPS; BLOOD GLUCOSE MONITORING TEST STRIPS Back to Search Results
Catalog Number 05942861001
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/04/2022
Event Type  malfunction  
Event Description
There was an allegation of questionable glucose results from accu-chek inform ii meter serial number (b)(4).At 10:56 am, the result was 401 mg/dl.At 10:59 am, the result was 276 mg/dl.At 9:04 pm, the result was 401 mg/dl.At 9:13 pm, the result was 235 mg/dl.No treatment was provided to the patient based on either questionable result of 401 mg/dl as the nurse stated the patient's baseline glucose results average in the 200's mg/dl and she did not believe the results of 401 mg/dl to be accurate.
 
Manufacturer Narrative
Qc was performed on (b)(6) 2022 and passed for both levels.The meter and strips have been requested for return.On a regular basis, accu-chek inform strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.The investigation is ongoing.
 
Manufacturer Narrative
No product was returned.No further investigation could be performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCU-CHEK INFORM II TEST STRIPS
Type of Device
BLOOD GLUCOSE MONITORING TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15698500
MDR Text Key303245198
Report Number1823260-2022-03432
Device Sequence Number1
Product Code LFR
UDI-Device Identifier00365702428102
UDI-Public00365702428102
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121679
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/31/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05942861001
Device Lot Number670103
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/15/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient SexMale
-
-