Catalog Number C1120
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that a cracked inspiratory filter connection and hole in circuit, indicated leaking when performing confirmatory water test.No patient injury was reported.
 
Manufacturer Narrative
This mdr was generated under protocol (b)(4), as a result of warning letter cms# 617147.No lot number was provided; therefore, device history record review could not be completed.A product sample was received for evaluation.Visual and functional testing were performed.The samples were visually inspected at a distance of 12? under normal lighting to received unit, in order to detect any damage on the unit.Two (2) samples have a tear.Functional test: one (1) sample was tested due to confirmation of leaking failure mode, sample was tested using (b)(4), the following gauge from of (b)(4) was used for the test.The circuit did not pass the leak test.A review of the manufacturing process for was conducted by quality engineer in order to verify that there are no situations or practices that could create the event.No discrepancies were found.A random sample of 32 units was taken from the manufacturing process, in order to perform a visual inspection to verify that all components were free of damage (scuffs, pinch marks, etc.), cracks, crazing, cuts or other workmanship defects that can affect assembly function or appearance.All the inspections and verifications are being performed by the production personnel and verified by the quality inspectors.The root cause of the reported issue was unable to be determined.Actions were taken to mitigate the reported issue: a corrective and preventive actions was opened, in order to determine the root cause and implement proper corrective actions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX GENERAL ANESTHESIA CIRCUITS
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15699512
MDR Text Key307074221
Report Number3012307300-2022-26510
Device Sequence Number1
Product Code BTO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/31/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberC1120
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2019
Was the Report Sent to FDA? No
Date Manufacturer Received09/29/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-