• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ULTRAFLEX TRACHEOBRONCHIAL; PROSTHESIS, TRACHEAL, EXPANDABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ULTRAFLEX TRACHEOBRONCHIAL; PROSTHESIS, TRACHEAL, EXPANDABLE Back to Search Results
Model Number M00564670
Device Problems Break (1069); Failure to Advance (2524); Activation, Positioning or Separation Problem (2906)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/30/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported to boston scientific corporation on (b)(6) 2022, that an ultraflex tracheobronchial uncovered proximal release stent was to be implanted to treat a tracheoesophageal fistula after radiotherapy due to esophageal cancer during an airway stent placement procedure performed on (b)(6) 2022.The patient's anatomy was dilated prior to stent placement.During the procedure, resistance was met, and the ultraflex tracheobronchial delivery system had difficulty crossing the stenosis.After multiple dilation, the device could still not cross the stenosis.It was noticed that the black stent deployment suture was loosened, and the ultraflex tracheobronchial stent partially opened.The physician adjusted the delivery system several times; however, the device could still not cross the stenosis.Subsequently, the black stent deployment suture broke.The ultraflex tracheobronchial stent was removed from the patient partially deployed, and a different device was implanted to complete the procedure.There were no patient complications as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAFLEX TRACHEOBRONCHIAL
Type of Device
PROSTHESIS, TRACHEAL, EXPANDABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15710709
MDR Text Key304626389
Report Number3005099803-2022-06370
Device Sequence Number1
Product Code JCT
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K012883
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 11/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/01/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/19/2023
Device Model NumberM00564670
Device Catalogue Number6467
Device Lot Number0026837069
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/12/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/19/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age60 YR
Patient SexMale
Patient Weight70 KG
-
-