It was reported to boston scientific corporation on (b)(6) 2022, that an ultraflex tracheobronchial uncovered proximal release stent was to be implanted to treat a tracheoesophageal fistula after radiotherapy due to esophageal cancer during an airway stent placement procedure performed on (b)(6) 2022.The patient's anatomy was dilated prior to stent placement.During the procedure, resistance was met, and the ultraflex tracheobronchial delivery system had difficulty crossing the stenosis.After multiple dilation, the device could still not cross the stenosis.It was noticed that the black stent deployment suture was loosened, and the ultraflex tracheobronchial stent partially opened.The physician adjusted the delivery system several times; however, the device could still not cross the stenosis.Subsequently, the black stent deployment suture broke.The ultraflex tracheobronchial stent was removed from the patient partially deployed, and a different device was implanted to complete the procedure.There were no patient complications as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|