Other, other text: h6: event problem and evaluation codes: updated.H3: device evaluated by manufacturer: updated h10: device evaluation: the device was returned for investigation.Visual inspection and functional test were performed.The customer reported problem was not related to any previous repair.Visual inspection found the device in good condition with no physical damage.The tamper seals and labels were intact.There was no evidence of the error recorded in the event history log.The customer reported problem was not verified/duplicated during functional testing.They were unable to determine the cause of the error.No fault was found.However, the silicone tubing connection to the manometer was replaced as a preventative measure.The root cause was undetermined/unknown.The previous service history has been reviewed and deemed unrelated to the current customer reported problem.This remediation mdr was generated under protocol b10010579, as a result of warning letter cms# (b)(6).
|