ST. JUDE MEDICAL, INC. AMPERE¿ RF ABLATION GENERATOR; EP GENERATOR Back to Search Results
Model Number H700489
Device Problem Failure to Deliver Energy (1211)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/26/2022
Event Type  malfunction  
Event Description
During the procedure, the generator did not ablate and the procedure was cancelled.A new cable, catheter, and rf grounding patch were used which did not resolve the issue and the procedure was cancelled.
 
Manufacturer Narrative
One ampere¿ rf ablation generator was received for evaluation.The returned generator was powered on, and the unit completed the post (power on self test) successfully and the screen display came up normally.Rf energy output was applied during ablation testing and no anomalies were identified throughout investigation as performed.The returned product functioned properly during the evaluation.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information provided to abbott and the investigation performed, the reported failure to deliver energy and subsequent cancellation remain unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPERE¿ RF ABLATION GENERATOR
Type of Device
EP GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key15713921
MDR Text Key305606728
Report Number2184149-2022-00261
Device Sequence Number1
Product Code OAD
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P110016
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/02/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberH700489
Device Lot Number7784332
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/28/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/14/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-