Model Number H700489 |
Device Problem
Failure to Deliver Energy (1211)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/26/2022 |
Event Type
malfunction
|
Event Description
|
During the procedure, the generator did not ablate and the procedure was cancelled.A new cable, catheter, and rf grounding patch were used which did not resolve the issue and the procedure was cancelled.
|
|
Manufacturer Narrative
|
One ampere¿ rf ablation generator was received for evaluation.The returned generator was powered on, and the unit completed the post (power on self test) successfully and the screen display came up normally.Rf energy output was applied during ablation testing and no anomalies were identified throughout investigation as performed.The returned product functioned properly during the evaluation.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information provided to abbott and the investigation performed, the reported failure to deliver energy and subsequent cancellation remain unknown.
|
|