SMITHS MEDICAL INTERNATIONAL, LTD. PNEUPAC VENTILATORS PARAPAC PLUS; VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)
|
Back to Search Results |
|
Model Number P310NUS |
Device Problem
Incorrect Measurement (1383)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Device evaluation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported the devices gage is not working.There has been no report of patient involvement or no observable clinical symptoms or a change in symptoms identified in the patient.
|
|
Manufacturer Narrative
|
H10: device evaluation: one ventilator device was returned to manufacturing for investigation.Visual inspection and functional testing were performed.Visual inspection of the returned device confirmed that the tamper seal was broken or removed upon receipt, the manometer gauge was out of spec, the battery cover was cracked, the tidal volume knob rubbed again the battery compartment and all small knobs were replaced with different knobs.Error codes from an event history log (ehl) are not available.After visual inspection, the customer reported issue was confirmed, the device manometer arrow was out of spec and stuck 18.The root cause is attributed to a faulty manometer.The manometer gauge was replaced to correct the issue.The previous service history has been reviewed and deemed unrelated to the current customer reported problem.This remediation mdr was generated under protocol b10010579, as a result of warning letter cms# 617147.
|
|