• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PNEUPAC VENTILATORS PARAPAC PLUS; VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PNEUPAC VENTILATORS PARAPAC PLUS; VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR) Back to Search Results
Catalog Number 530A1167NPL
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Udi is unknown.No information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
I was reported that there was a leak inside the ventilator.There was no patient involved reported.
 
Manufacturer Narrative
Other text: h6: event problem and evaluation codes: updated.H10: device evaluation: the device was returned for investigation.A visual inspection and functional test were performed.Visual inspection found the device in good condition.The customer stated problem was duplicated.Leaking demand valve assembly was found, the demand detector assembly was replaced.The cause of the reported problem could not be determined.A dhr (device history review) was performed and no problems or issues were identified during this dhr review.No previous service history for this device was located in the database.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PNEUPAC VENTILATORS PARAPAC PLUS
Type of Device
VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
bramingham business park, enterprise way
luton, bedforshire LU3 4 BU
UK  LU3 4BU
Manufacturer (Section G)
NULL
bramingham business park, enterprise way
luton, bedforshire LU3 4 BU
UK   LU3 4BU
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15716288
MDR Text Key307464225
Report Number3012307300-2022-26634
Device Sequence Number1
Product Code BTL
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
K123957
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/02/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number530A1167NPL
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received11/14/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/27/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-