Model Number CI-1601-05 |
Device Problems
Energy Output Problem (1431); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Undesired Nerve Stimulation (1980); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Type
Injury
|
Event Description
|
The recipient is reportedly experiencing poor performance, and non-auditory sensations.Device testing was attempted but was discontinued due to the recipient being unable to tolerate loudness.Programming adjustments were made; however, the issue did not resolve.Revision surgery is under consideration.
|
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
A ct scan confirmed proper device placement.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was sliced.This is believed to have occurred during revision surgery.The photographic imaging inspection revealed a broken electrode wire.This is believed to have occurred during revision surgery.System lock was verified.The condition of the electrode prevented an electrical test from being performed.The device passed the electrical and mechanical tests performed.The reported complaint of impedance issues could not be verified during this analysis, which was limited in some respects due to the electrode being sliced prior to receipt.The device passed all the of the tests performed.This version of the hires ultra device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|