DAVOL INC., SUB. C.R. BARD, INC. -1213643 HYDRO-SURG PLUS LAPAROSCOPIC IRRIGATOR; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Catalog Number 0026870
Device Problems Thermal Decomposition of Device (1071); Corroded (1131); Fluid/Blood Leak (1250); Smoking (1585)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/07/2022
Event Type  malfunction  
Manufacturer Narrative
As reported, the bard/davol hydrosurg laparoscopic irrigation system started to smoke at the end of procedure.The subject device has been returned.Evaluation identifies the ingress of saline fluid beyond the motor mount and into the unit.Condition resulted in a thermal event (short circuit) related to saline in contact with the circuit board.Based on our evaluation and the condition of the device, the root cause is identified as manufacturing related.The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
As reported per mi case# (b)(4): on (b)(6) 2022 a bard/davol hydrosurg irrigator device started to smoke at the end of the procedure.It was reported that the device was immediately removed from the room and removed the battery.As reported, they were able to complete the procedure and no patient harm was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HYDRO-SURG PLUS LAPAROSCOPIC IRRIGATOR
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC. -1213643
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
DAVOL SURGICAL INNOVATIONS -9616067
ave. roberto fierro #6408
parque industrial aeropuerto
cd. juarez, chih s.a. de c.v. 32690
MX   32690
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
8005566756
MDR Report Key15721911
MDR Text Key303024579
Report Number1213643-2022-00704
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier00801741000362
UDI-Public(01)00801741000362
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K961492
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/03/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0026870
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/01/2022
Date Manufacturer Received10/12/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-