Visual and functional analysis were performed on the returned device.The reported signal calibration failure was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the product investigation/complaint history of the reported lot did not indicate a lot specific quality issue.The investigation determined the reported signal calibration failure was likely due to operational context.Return analysis noted multiple fibers on the proximal end of the contact members of the proximal male connector area.In this case, it is likely that the transmitter was removed from the wire during preparation and wiped down with a cloth (or the like), which resulted in the inability to connect the device.The fibers likely caused insufficient electrical contact surface or interference between the guidewire and pwx transmitter, and subsequently lead to the reported/observed signal calibration failure.Once the fibers were removed from the contact members the device was able to be successfully connected, as expected without error.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
It was reported that that the pressurewire x - wireless device could not be calibrated.The device failed to connect to the system and the lighting pattern was solid yellow.The device was confirmed to be fully inserted.The system was shut down and restarted.However, the same issue persisted.Therefore, the device was not used in the patient and the procedure was completed with another device.There was no patient involvement and no clinically significant delay in the procedure.Device return analysis revealed that the pressurewire was then removed from the transmitter and there were multiple fibers observed on the proximal end of the proximal male connector area.No additional information was provided.
|