SMITH & NEPHEW, INC. INTEGRAL BIPOLAR CUP SET 44MM; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Catalog Number 11044
Device Problem Migration or Expulsion of Device (1395)
Patient Problems Pain (1994); Inadequate Osseointegration (2646)
Event Date 10/05/2022
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that, after a thr surgery was performed on (b)(6) 2001, the patient experienced pain and migration of the implant.This adverse event was treated by a revision surgery on (b)(6) 2022, in which an integral bipolar cup set 44mm was explanted.Patient's current health status is unknown.
 
Manufacturer Narrative
The device was not returned for evaluation; therefore, a device analysis could not be performed.The clinical/medical investigation concluded that, as of the date of this medical investigation, the requested clinical documentation has not been provided; therefore, there were no clinical factors found which would have contributed to the event.Procedure was completed with a back-up device.No injury to the patient was reported.Device batch number was not provided, thus, an evaluation of the manufacturing records could not be performed.A review of complaint history for the previous 12 months did not reveal similar events for the listed device.A review of the instructions for use documents for total hip systems revealed that implant migration or subsidence that has resulted in revision surgery and has occurred in conjunction with compaction grafting procedures can occur usually as a result of insufficient graft material, improper cement techniques, and/or varus stem alignment has been identified as an adverse events in primary and revision surgery.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include patient anatomy and/or traumatic injury.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTEGRAL BIPOLAR CUP SET 44MM
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15728683
MDR Text Key303028353
Report Number1020279-2022-04666
Device Sequence Number1
Product Code JDG
UDI-Device Identifier00885556007679
UDI-Public885556007679
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/04/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number11044
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/16/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age70 YR
Patient SexFemale
-
-