Model Number 28-95-80 |
Device Problems
Decreased Pump Speed (1500); Increased Pump Speed (1501)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/05/2022 |
Event Type
malfunction
|
Event Description
|
Livanova received report of a s5 mast roller pump encoder failure visible and no error message was shown on the control display module.The issue was identified during maintenance and there was no patient involvement.
|
|
Manufacturer Narrative
|
There was no patient involvement.Livanova deutschland manufactures the s5 mast roller pump.The incident occurred in (b)(6).The technician in charge replaced the encoder and the problem was solved.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Manufacturer Narrative
|
H10: based on the investigation performed for similar events, the most likely root cause of the reported event is the defective shaft angle encoder.
|
|
Event Description
|
See initial report.
|
|