SMITHS MEDICAL ASD, INC. BIVONA NEONATAL AND PEDIATRIC TTS (TIGHT-TO-SHAFT); TUBE TRACHEOSTOMY AND TUBE CUFF
|
Back to Search Results |
|
Model Number 67SP045 |
Device Problem
Nonstandard Device (1420)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/14/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported that the cuff was checked before patient use, it was not symmetrical, and applied pressure to the patients trachea.No medical intervention was required.No patient injury was reported.
|
|
Manufacturer Narrative
|
A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.Three samples were received for evaluation.Samples were filled with 5cc of air to see if there was any functional problem, then the cuff was measured for symmetry.Cuff was inflated symmetrically for three samples; complaint was not confirmed.Devices passed all functional testing.No problem found.Root cause cannot be determined as the sample was successfully tested and no discrepancies were detected.This issue will be monitored for any increase in occurrence.If the occurrence of this issue increases significantly, additional investigative action will be taken accordingly.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
|
|