Model Number MU-1 |
Device Problem
Power Problem (3010)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/29/2022 |
Event Type
malfunction
|
Event Description
|
The customer reported to olympus, the maintenance unit back power receptacle wouldn¿t work.The event was found during reprocessing.There was no report of patient harm associated with this event.The subject device was returned to olympus for evaluation.During inspection and testing, the allegation was confirmed.The alternating current (ac) inlet at the rear was damaged and pushed inward, resulting in the inability to connect power cord.This report is being submitted for the malfunction found at estimation (damaged ac inlet).
|
|
Manufacturer Narrative
|
During inspection and testing, the customer's allegation was confirmed.In addition, air pressure fluctuated due to worn out air joint unit and connector socket.The rear panel had a deformity.There were 4 suction feet at the bottom with a weak suction force.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause could not be determined.It was confirmed that the power inlet was damaged but the cause of the damage could not be identified.Olympus will continue to monitor field performance for this device.
|
|