TERUMO PHILIPPINES CORPORATION MCKESSON PREVENT ® HT SAFETY HYPODERMIC NEEDLE; SYRINGE, ANTISTICK Back to Search Results
Model Number N/A
Device Problem Activation Problem (4042)
Patient Problem Needle Stick/Puncture (2462)
Event Date 10/05/2022
Event Type  malfunction  
Event Description
The user facility reported that a staff member experienced challenges with activating the safety with the involved needle.Additional information was received on 07 october 2022: it was reported that the challenge that was faced with activation was the safety activated however it would not fully cover the needle.After a staff member activated the needle, when placing the needle in the sharp's container, their finger was scratched on the needle.The staff member did activate the needle against a hard surface as indicated in ifu.Hiv testing was performed after the needle to ensure the wellbeing of the staff member.The estimated blood loss was less than 250cc.The staff member is stable/ok.
 
Manufacturer Narrative
Implanted date: device was not implanted.Explanted date: device was not explanted.Occupation: msn, rn.The actual device has not been returned for evaluation.The root cause of the complaint could not be identified to be related to our production or process.Retention samples were visually confirmed free from defects that will affect the activation of the safety sheath and passed the evaluation for sheath activation and deactivation.Also, no irregularity was encountered during the simulation of manual sheath activation that may lead to the complaint.We have a series of visual in-process inspections to detect an abnormality on the sheath that may lead to a problem during sheath activation.The molding condition of the components critical to the safety activation of the product is routinely checked to assure that no defects will be encountered that will lead to sheath activation problems.Similarly, the assembly status of the safety needle such as sheath collar fitting and damaged parts that will affect product function is being confirmed.Lot history file revealed no related nonconformity or irregularity that will lead to the complaint.Prior to shipment, qc conducts outgoing visual, sensory, and functional inspections to assure lots are of good quality.Therefore, we advise to follow the instructions for use (ifu) for the proper usage of the sg2 needle indicated on the unit box in which warnings to prevent needle stick, cautions, and precautions are also included.The investigation is currently ongoing.A follow-up report will be submitted once the investigation is complete.Terumo medical products (tmp) (importer) registration no.(b)(4) is submitting this report on behalf of terumo (philippines) corporation (manufacturer) registration no.(b)(4).
 
Manufacturer Narrative
This report is being submitted as follow-up no.1 to provide the unused sample return date in section d9, to update section h3, and to provide the completed investigation results.The actual sample was not available for evaluation hence we cannot provide the details of its actual condition.Instead, seventeen (17) pieces of unused same lot samples were received.The samples were visually confirmed free from defects that will affect activation of safety sheath.There was no further confirmation on the received same lot sample since it has the same condition as the retention sample.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCKESSON PREVENT ® HT SAFETY HYPODERMIC NEEDLE
Type of Device
SYRINGE, ANTISTICK
Manufacturer (Section D)
TERUMO PHILIPPINES CORPORATION
124 east main avenue
laguna technopark
binan,
RP 
Manufacturer (Section G)
TERUMO PHILIPPINES CORPORATION
124 east main avenue
laguna technopark
binan, laguna,,
RP  
Manufacturer Contact
gina digioia
265 davidson ave
suite 320
somerset, NJ 08873
6402040886
MDR Report Key15741021
MDR Text Key305902661
Report Number3003902955-2022-00048
Device Sequence Number1
Product Code MEG
UDI-Device Identifier30612479178855
UDI-Public30612479178855
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K051865
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 11/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/05/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number102-N251S
Device Lot Number210807B
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/04/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/07/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age37 YR
Patient SexFemale
Patient Weight91 KG
Patient EthnicityHispanic
Patient RaceWhite
-
-