Lot Number 0028860834 |
Device Problem
Component Missing (2306)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/18/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that a labelling issue occurred.A 3.00 x 32mm synergy xd drug-eluting stent was selected for use.However, when the packaging was opened, it was found that the stent was unmarked making it unusable.Another of the same device was used and completed the procedure.There were no patient complications reported.
|
|
Manufacturer Narrative
|
Device evaluated by mfr.: synergy xd mr ous 3.00 x 32mm; stent delivery system, catheter was returned for analysis.The device packaging was not received.A visual examination of the stent found stent damage.Stent damage with struts near the proximal region were bunched.The undamaged stent outer diameter was measured, and the result was within max crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip identified no damage.A visual and microscopic and tactile examination of the hypotube shaft found no kinks or damages.A visual and tactile examination of the outer and mid-shaft section and a visual examination of the inner lumen found no issues along the shaft polymer extrusion.An examination of the hub confirmed that the product details were clearly printed on the hub.
|
|
Event Description
|
It was reported that a labelling issue occurred.A 3.00 x 32mm synergy xd drug-eluting stent was selected for use.However, when the packaging was opened, it was found that the stent was unmarked making it unusable.Another of the same device was used and completed the procedure.There were no patient complications reported.
|
|
Search Alerts/Recalls
|