Model Number N/A |
Device Problems
Electrical /Electronic Property Problem (1198); Mechanical Problem (1384); Gas/Air Leak (2946); Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/11/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Report source - report source - foreign: (b)(4).The product has been returned to zimmer biomet, and the investigation is in process.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Investigation incomplete.
|
|
Event Description
|
It was reported that outside of surgery the unit doesn't work with or without the hose.The device stopped working after being used for many years.The compressed air hose was leaking, but after repair, the dermatome still did not work.No patient involvement or harm.Attempts have been made, but no further event information available at the time of this report.
|
|
Event Description
|
It was reported that outside of surgery the unit doesn't work with or without the hose.The device stopped working after being used for many years.The device was leaking air.No patient involvement or harm.Due diligence is complete and no additional information is available.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under: (b)(4).This medwatch is being filed to relay additional information.Review of the most recent repair record determined the unit was not working properly as the motor ran below specifications and the swivel was loose.The needle bearings and reciprocating arm were found to be heavily corroded.The unit was also out of calibration and the control bar was out of position.The motor, sleeve, swivel, needle bearing, and reciprocating arm were replaced and the unit and control bar were recalibrated to resolve the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.It was noted that the device has not been regularly returned for annual pm.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|