• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME; DERMATOME HANDPIECE, PNEUMATIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME; DERMATOME HANDPIECE, PNEUMATIC Back to Search Results
Model Number N/A
Device Problems Electrical /Electronic Property Problem (1198); Mechanical Problem (1384); Gas/Air Leak (2946); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/11/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source - report source - foreign: (b)(4).The product has been returned to zimmer biomet, and the investigation is in process.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Investigation incomplete.
 
Event Description
It was reported that outside of surgery the unit doesn't work with or without the hose.The device stopped working after being used for many years.The compressed air hose was leaking, but after repair, the dermatome still did not work.No patient involvement or harm.Attempts have been made, but no further event information available at the time of this report.
 
Event Description
It was reported that outside of surgery the unit doesn't work with or without the hose.The device stopped working after being used for many years.The device was leaking air.No patient involvement or harm.Due diligence is complete and no additional information is available.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under: (b)(4).This medwatch is being filed to relay additional information.Review of the most recent repair record determined the unit was not working properly as the motor ran below specifications and the swivel was loose.The needle bearings and reciprocating arm were found to be heavily corroded.The unit was also out of calibration and the control bar was out of position.The motor, sleeve, swivel, needle bearing, and reciprocating arm were replaced and the unit and control bar were recalibrated to resolve the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.It was noted that the device has not been regularly returned for annual pm.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME HANDPIECE, PNEUMATIC
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key15751260
MDR Text Key303559112
Report Number0001526350-2022-01096
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/08/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot Number61919687
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/21/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/15/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/18/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE H10.
-
-