• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. SINGLE USE LIGATING DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. SINGLE USE LIGATING DEVICE Back to Search Results
Model Number HX-400U-30
Device Problems Break (1069); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/14/2022
Event Type  malfunction  
Manufacturer Narrative
The device is not returned, as such a definitive root cause of the reported complaint cannot be determined at this time.This event is under investigation.A supplemental report will be submitted upon completion of the investigation or upon receiving additional information.
 
Event Description
As reported for this event by the customer, during a therapeutic polypectomy procedure, the device handle was difficult to operate, and the polyp was not well ligated at the time of litigating the polyp.When releasing, the device handle was stiff and could not be released.The device handle broke and the loop was loosened before release.The procedure was continued and completed.There is no harm or adverse impact to the patient.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the approved final investigation.The device was manufactured in december 2021.However, the specific date is unknown because the device was not returned.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, and since the device was discarded, the definitive root cause of the loop issue could not be determined.The loop detachment may have been caused by the loop being detached from the hook while the coil sheath was being retracted into the tube sheath, or the loop was hung on to the tissue and it was positionally fixed at the distal end of the tube, rendering the loop unable to release.The tube sheath may have been pulled toward the proximate side while the hook was extended from the distal end of the coil sheath, which may have caused the loop to move with the distal end of the tube sheath toward the coil sheath and got caught between the hook and the coil sheath.The instruction manual identifies the following verbiage, which may have prevented the phenomenon: ¿do not strike or crush the coil sheath during operation.Doing so can damage the distal end of the coil sheath, which could make it impossible to detach the loop after ligation.In this case, refer to section 12, ¿emergency treatment¿ and as shown ¿equipment to be used in an emergency¿ on page 3 in this manual.Do not remove the loop from the hook while the coil sheath is not extended from the tube sheath.Otherwise, the loop may be tangled with the hook and become impossible to be removed.In this case, refer to section 12, ¿emergency treatment¿ and as shown ¿equipment to be used in an emergency¿ on page 3 in this manual.Do not hold the loop with the distal end of the tube sheath while the loop is surrounding the tissue.Otherwise, when the tissue is ligated, the loop may be detached from the hook in the tube sheath and tangled with the hook.That may make the loop impossible to be removed.In this case, refer to section 12, ¿emergency treatment¿ and as shown ¿equipment to be used in an emergency¿ on page 3 in this manual.Never use excessive force to operate the instrument.This could damage the instrument.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE LIGATING DEVICE
Type of Device
SINGLE USE LIGATING DEVICE
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer Contact
masaharu hirose
2-248-1 okkonoki
kuroishi-shi, aomori 036-0-357
JA   036-0357
426422891
MDR Report Key15753715
MDR Text Key307110202
Report Number9614641-2022-00575
Device Sequence Number1
Product Code FHN
UDI-Device Identifier04953170194641
UDI-Public04953170194641
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/08/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHX-400U-30
Device Lot Number1ZK
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/24/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/01/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-