As reported for this event by the customer, during a therapeutic polypectomy procedure, the device handle was difficult to operate, and the polyp was not well ligated at the time of litigating the polyp.When releasing, the device handle was stiff and could not be released.The device handle broke and the loop was loosened before release.The procedure was continued and completed.There is no harm or adverse impact to the patient.
|
This report is being supplemented to provide additional information based on the approved final investigation.The device was manufactured in december 2021.However, the specific date is unknown because the device was not returned.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, and since the device was discarded, the definitive root cause of the loop issue could not be determined.The loop detachment may have been caused by the loop being detached from the hook while the coil sheath was being retracted into the tube sheath, or the loop was hung on to the tissue and it was positionally fixed at the distal end of the tube, rendering the loop unable to release.The tube sheath may have been pulled toward the proximate side while the hook was extended from the distal end of the coil sheath, which may have caused the loop to move with the distal end of the tube sheath toward the coil sheath and got caught between the hook and the coil sheath.The instruction manual identifies the following verbiage, which may have prevented the phenomenon: ¿do not strike or crush the coil sheath during operation.Doing so can damage the distal end of the coil sheath, which could make it impossible to detach the loop after ligation.In this case, refer to section 12, ¿emergency treatment¿ and as shown ¿equipment to be used in an emergency¿ on page 3 in this manual.Do not remove the loop from the hook while the coil sheath is not extended from the tube sheath.Otherwise, the loop may be tangled with the hook and become impossible to be removed.In this case, refer to section 12, ¿emergency treatment¿ and as shown ¿equipment to be used in an emergency¿ on page 3 in this manual.Do not hold the loop with the distal end of the tube sheath while the loop is surrounding the tissue.Otherwise, when the tissue is ligated, the loop may be detached from the hook in the tube sheath and tangled with the hook.That may make the loop impossible to be removed.In this case, refer to section 12, ¿emergency treatment¿ and as shown ¿equipment to be used in an emergency¿ on page 3 in this manual.Never use excessive force to operate the instrument.This could damage the instrument.¿ olympus will continue to monitor field performance for this device.
|