• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COLOPLAST A/S TITAN TOUCH; INFLATABLE PENILE PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COLOPLAST A/S TITAN TOUCH; INFLATABLE PENILE PROSTHESIS Back to Search Results
Model Number ES29222400
Device Problems Failure to Cycle (1142); Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  Injury  
Event Description
According to available information, this device required replacement due to a tubing break.There appeared to be a tubing break near the pump.It was also noted that the cylinders were not cycling.No other adverse patient effects were reported.
 
Manufacturer Narrative
The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release and no trends were noted.Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information or conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TITAN TOUCH
Type of Device
INFLATABLE PENILE PROSTHESIS
Manufacturer (Section D)
COLOPLAST A/S
holtedam 1
humlebæk 3050
DA  3050
Manufacturer (Section G)
COLOPLAST MANUFACTURING US, LLC
1601 west river road north
minneapolis MN 55411
Manufacturer Contact
usbes brian e schmidt
1601 west river road north
minneapolis, MN 55411
8007880293
MDR Report Key15759028
MDR Text Key303308794
Report Number2125050-2022-01220
Device Sequence Number1
Product Code FHW
UDI-Device Identifier05708932539203
UDI-Public05708932539203
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P000006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 11/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/09/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/12/2019
Device Model NumberES29222400
Device Catalogue NumberES2922
Device Lot Number4451807
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/01/2022
Date Device Manufactured01/06/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age86 YR
Patient SexMale
-
-