It was reported that there was an issue with pl574t - challenger ti-p sm-ligat.Clips 12 cartr.According to the complaint description, the clip cartridge jumped out and fragmented in the patient, at the time of installation.An additional medical intervention was required.The fragment was retrieved and there was a short surgical delay.The malfunction is filed under aag reference (b)(4).
|
Updated b3 date of the event.Updated d9 - no product returned.Updated h6 codes.Investigation results: as of the date of this report the complaint product was not provided for investigation.Therefore, a thorough investigation is not possible.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.In the event that the complaint product will be provided for investigation in the future, an update of this report will be provided unsolicited.Based upon the investigations results a capa is not necessary.
|