A portion of the csl was received at cvrx for analysis.The returned csl did not include the electrode section.The csl was tightly tangled, and the inner coil beyond the suture tab was stretched, broken, and exposed with no insulation.The lead was unable to be functionally tested as the electrode was not returned.No manufacturing defects were observed.The root cause of the event was determined to be related to the patient twiddling their system.The device history record and sterilization record for these device serial numbers have been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The devices met material, assembly, and quality control requirements.Cvrx id# (b)(4).
|
A barostim system was implanted on (b)(6) 2022.On (b)(6) 2022, the patient had their second follow-up appointment, the lead impedance was low, and the compliance failed.Additional impedance tests were performed, but the issue persisted.The lead impedance trend showed that the impedance dropped low and remained low on (b)(6) 2022 in the evening.The patient reported experiencing a sharp neck pain around that time, which was followed by intermittent ipg movement.A chest and neck x-ray was performed, the lead was balled up near the ipg, and the ipg had flipped.A revision was performed on (b)(6) 2022.It was found that the ipg sutures were still in place on the ipg but had pulled out of the tissue.The device had been twiddled, and there were many tension loops in the csl.The csl had fractured.The csl was replaced, and the ipg was reimplanted with the sutures being placed deeper in the tissue.The patient was doing well following the procedure.
|