• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (IRVINE) XEN 45 GTS; IMPLANT, EYE VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (IRVINE) XEN 45 GTS; IMPLANT, EYE VALVE Back to Search Results
Catalog Number 5507-001
Device Problem Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/27/2022
Event Type  malfunction  
Manufacturer Narrative
Further information regarding event, product, or patient details has been requested.No additional information is available at this time.A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.
 
Event Description
Healthcare professional reported a xen®45 gts event described as the slider got stuck and implant broke during implantation in left eye.Surgery was completed with second xen®45 gts device.No eye injury noted.
 
Event Description
Healthcare professional reported a xen®45 gts event described as the slider got stuck and implant broke during implantation in left eye.Surgery was completed with second xen®45 gts device.No eye injury noted.
 
Manufacturer Narrative
Device analysis: the needle cover was attached.The retention plug and cam lock were each detached and not returned.The xen gel stent was not returned.A visual evaluation of the xen injector was performed.The slider of the injector was not in the start position and was significantly advanced, the needle was fully retracted, and the bevel was in the center position.The slider knob was able to slide the full distance of the travel length in each of the three bevel selector positions.The functional testing showed that the pusher rod which pushes the gel stent out of the needle performed correctly, and the needle moved in tandem with the slider knob.The needle was fully extended as the slider knob was at the start position, and the needle retracted properly as the slider knob was advanced to the end of the travel distance.Since the slider knob was able to slide the full distance of the travel length of each of the three bevel selector positions and since resistance was not observed, the expected condition was met.The reported complaint of blocked slider is not confirmed since the slider knob was able to slide the full distance of the travel length of each of the three bevel selector positions and since resistance was not observed during functional testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XEN 45 GTS
Type of Device
IMPLANT, EYE VALVE
Manufacturer (Section D)
ALLERGAN (IRVINE)
2525 dupont drive
irvine CA 92612
Manufacturer (Section G)
ALLERGAN (IRVINE)
2525 dupont drive
irvine CA 92612
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key15777357
MDR Text Key306957756
Report Number3011299751-2022-00187
Device Sequence Number1
Product Code KYF
UDI-Device Identifier01240000000405
UDI-Public1240000000405
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K161457
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/10/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number5507-001
Device Lot Number63990
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/30/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexPrefer Not To Disclose
-
-