Model Number 1225250
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Wound Dehiscence (1154); Unspecified Infection (1930)
Event Date 07/12/2022
Event Type  Injury  
Manufacturer Narrative
This complaint has been created from a clinical study review so no product returned for evaluation and no lab reports or images were provided so the complaint cannot be confirmed.A definitive root cause cannot be determined though post operative infection is a known complication of spinal surgery.Nuvasive non-sterile implants and instrumentation are cleaned and processed by the end user and sterilization records were not provided.No additional investigation can be completed.Labeling review: ".Contraindications: contraindications include, but are not limited to: infection, local to the operative site.Signs of local inflammation." ".Potential adverse events and complications: as with any major surgical procedures, there are risks involved in spinal/orthopedic surgery.Infrequent operative and postoperative complications that may result in the need for additional surgeries include: early or late infection.Potential risks identified with the use of this system, which may require additional surgery, include: infection." ".Warnings, cautions and precautions: care should be taken to ensure that all components are ideally fixated prior to closure." ".Pre-operative warnings: patient condition and/or predispositions such as those addressed in the aforementioned contraindications should be avoided.Care should be used during surgical procedures to prevent damage to the devices and injury to the patient." ".Cleaning and decontamination: all non-sterile instruments must first be thoroughly cleaned using the validated methods prescribed in the nuvasive cleaning and sterilization instructions doc 9400896 before sterilization and introduction into a sterile surgical field.Contaminated instruments should be wiped clean of visible soil at the point of use, prior to transfer to a central processing unit for cleaning and sterilization.The validated cleaning methods include both manual and automated cleaning.Visually inspect the instruments following performance of the cleaning instructions to ensure there is no visual contamination of the instruments prior to proceeding with sterilization.If possible contamination is present at visual inspection, repeat the cleaning steps.Contaminated instruments should not be used, and should be returned to nuvasive.Contact your local representative or nuvasive directly for any additional information related to cleaning of nuvasive surgical instruments." ".Sterilization: all non-sterile instruments are sterilizable by steam autoclave using standard hospital practices, in addition to nuvasive¿s validated parameters.In a properly functioning and calibrated steam sterilizer, effective sterilization may be achieved using the parameters prescribed in the nuvasive cleaning and sterilization instructions doc 9400896." device in-situ.
 
Event Description
It was reported on a clinical study that on (b)(6) 2022 a patient underwent an extreme lateral interbody fusion at l4/5.On (b)(6) 2022 the patient had an er visit during which she had wound vac applied for infected lower posterior back wound.On (b)(6) 2022 the patient supposedly underwent incision and drainage for the wound dehiscence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE MODULUS XLIF INTERBODY SYSTEM
Type of Device
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
Manufacturer (Section D)
NUVASIVE, INCORPORATED
7475 lusk boulevard
san diego CA 92121
Manufacturer (Section G)
NUVASIVE, INCORPORATED
7475 lusk boulevard
CA 92121
Manufacturer Contact
geoff gannon
7475 lusk boulevard
san diego, CA 92121
MDR Report Key15784439
MDR Text Key303572171
Report Number2031966-2022-00251
Device Sequence Number1
Product Code MAX
UDI-Device Identifier00887517726919
UDI-Public887517726919
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201692
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Study,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/11/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number1225250
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/17/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age61 YR
Patient SexFemale
-
-