In addition to the findings reported, the balloon water tube was damaged causing water tightness to be lost.Due to clogging of nozzle, water removal ability did not meet specification.The bending angle in up, down and left direction was insufficient due to the elongation of the angle wire.The play of the up/down angle knob was out of specification due to the elongation of the angle wire.The bending section cover was dirty.The control unit was scratched.The acoustic lens was scratched.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation, or if additional information is received.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Please see the updates in sections e3, h6, and h10.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely due to insufficient cleaning (handling problems).The following information is stated in the instructions for use (ifu): ¿caution: ·after using the endoscope reprocess it according to the instructions given in chapter 7, ¿cleaning, disinfection, and sterilization procedures¿.Using improperly or incompletely reprocessed, the endoscope¿s distal end damage may result.Chapter 7 cleaning, disinfection, and sterilization procedures.Warning: ·all channels of the endoscope, including the elevator wire channel and balloon channel, must be cleaned and high-level disinfected or sterilized during every reprocessing cycle, even if the channels were not used during the previous patient procedure.Otherwise, insufficient cleaning and disinfection or sterilization of the endoscope may pose an infection control risk to the patient and/or operators performing the next procedure with the endoscope.¿ olympus will continue to monitor field performance for this device.
|