Model Number 00-3020
Device Problem Failure to Deliver Energy (1211)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/19/2022
Event Type  Injury  
Event Description
Customer reported tempus ls device pacer was not working.A user report was received related to a reported product problem which is currently being investigated.At the time of reporting, the device has not yet been returned for investigation.Further updates will be provided when the device is received, and the investigation is in progress.
 
Event Description
This report is based on information provided by philips remote service engineer and schiller investigation team and has been investigated by the philips complaint handling team.Philips received a complaint on the tempus ls indicating that the pacer is not working.The customer was performing annual preventative maintenance using a fluke 7000dp when the pacer would not increase and then the increase option disappeared.No patient care was being done.
 
Manufacturer Narrative
Based on the evidence available the reported problem was unconfirmed.The investigations involved a "good faith effort" to obtain information and the device.Despite all the efforts, sag was not able to obtain them.Therefore, no investigation could be done.This assessment will be updated, if new information are made available.The data entered in this complaint record will be utilized for product quality and safety improvements per the post market surveillance and risk management processes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEMPUS LS-MANUAL
Type of Device
LOW ENERGY DEFIBRILLATOR
Manufacturer (Section D)
SCHILLER AG
altgasse 68
baar 6341
SZ  6341
Manufacturer (Section G)
SCHILLER AG
altgasse 68
baar 6341
SZ   6341
Manufacturer Contact
rekha rv
ascent 1, aerospace centre
aerospace boulevard
farnborough GU14 -6XW
UK   GU14 6XW
MDR Report Key15797029
MDR Text Key303633721
Report Number3003832357-2022-00049
Device Sequence Number1
Product Code LDD
UDI-Device Identifier07613365002737
UDI-Public7613365002737
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200849
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/15/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number00-3020
Device Catalogue Number989706001681
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received12/22/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-