During an atrial fibrillation ablation 3d mapping procedure, signal interference was identified and the catheter was difficult to position in the patient.Upon removal from the patent, the electrode was noted to be separated from the shaft.Another device was used to continue this procedure and consequences to the patient.No additional information is available.
|
One 4mm tip, medium sweep, safire ablation catheter was received for evaluation.The distal tip was noted to be separated from the distal tip.Bends were noted in the shaft at 1.5¿, 2.6¿ and 5.7¿ proximal to the distal tip.Functional testing revealed the catheter was able to deflect in both directions when actuating the steering mechanism.The shaft did not deflect into the correct shape in either direction and no longer met specifications due to the bends in the shaft.Electrodes 1-4 and the thermistor/thermocouple met specifications during electrical testing, temperature testing could not be completed due to aforementioned damage.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported displaced electrode could not be conclusively determined.
|