BAXTER HEALTHCARE CORPORATION MINICAP TRANSFER SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
|
Back to Search Results |
|
Catalog Number 5C4483 |
Device Problems
Fluid/Blood Leak (1250); Noise, Audible (3273)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/20/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that a leak was observed while a peritoneal dialysis (pd) transfer set was connected to a minicap.This occurred when the cap was placed on the end of the transfer set during continuous ambulatory pd therapy.The nurse heard a cracking sound and noticed a fluid leak.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
H10: the device was received for evaluation.A visual inspection with the naked eye identified a broken occluder foot on the main body (light blue) beneath the white twist clamp.Functional testing including clear passage test was performed with no issues noted.Leak testing was performed with the returned mini cap attached to the female connector; no external leaks were noted.Clamp function testing failed due to a leak through the set due to the broken occluder foot.The reported condition was verified.The direct cause of the event was undetermined; however, the damaged set is consistent with damage caused by exposure to chemical agents such as foreign solvents, dyes, or cleaning agents that damage the transfer set materials.A nonconformance has been opened to address this issue.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Search Alerts/Recalls
|
|
|