APPLIED MEDICAL RESOURCES C2A15, GELPOINT V-PATH PLATFORM, 7CM; CULDOSCOPE (AND ACCESSORIES) Back to Search Results
Model Number C2A15
Device Problem Improper or Incorrect Procedure or Method (2017)
Patient Problem Laceration(s) (1946)
Event Date 10/19/2022
Event Type  Injury  
Manufacturer Narrative
No product is being returned to applied medical for evaluation.A follow-up report will be provided upon completion of investigation.
 
Event Description
Procedure performed: tubal ligation.Event description: c2a15 introducer induced trauma to patient's bowel when creating colpotomy.The surgeon sutured the trauma, and the case was completed using the same device as normal.No patient injury occurred.Product is not returning.Intervention: the surgeon sutured the trauma, and the case was completed using the same device as normal.Patient status: c2a15 introducer induced trauma to patient's bowel.No patient injury occurred.
 
Manufacturer Narrative
The event unit was not returned to applied medical for evaluation.In the absence of the event unit, it is difficult to determine the root cause of the incident.According to a subject matter expert, a surgeon who is familiar with the device, it is unlikely that the event unit caused the bowel injury due to the bluntness of the introducer.Applied medical has reviewed the details surrounding the event and is unable to determine the exact root cause of the injury.The probability and criticality of harm resulting from this failure have been evaluated and were found to be at an acceptable level.
 
Event Description
Procedure performed: tubal ligation.Event description: c2a15 introducer induced trauma to patient's bowel when creating colpotomy.The surgeon sutured the trauma, and the case was completed using the same device as normal.No patient injury occurred.Product is not returning.Intervention: the surgeon sutured the trauma, and the case was completed using the same device as normal.Patient status: c2a15 introducer induced trauma to patient's bowel.No patient injury occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
C2A15, GELPOINT V-PATH PLATFORM, 7CM
Type of Device
CULDOSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
Manufacturer Contact
aaron fulcher
22872 avenida empresa
rancho santa margarita, CA 92688
9497135765
MDR Report Key15806445
MDR Text Key303740977
Report Number2027111-2022-00823
Device Sequence Number1
Product Code HEW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191294
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/16/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberC2A15
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received10/19/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-