Model Number AB-5100R |
Device Problems
Misconnection (1399); Intermittent Communication Failure (4038)
|
Patient Problems
Failure of Implant (1924); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 11/21/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Event Description
|
The recipient is reportedly experiencing loss of lock.External equipment was changed, however the issue did not resolve.Revision surgery will be scheduled.
|
|
Manufacturer Narrative
|
The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed.In addition, inspection revealed damage to the epoxy header and braze.These anomalies were believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock could not be obtained at any spacing.The electrode condition prevented an electrical test from being performed.The device passed some of the electrical tests performed.The device passed the mechanical test performed.It is believed that the failure of this implantable cochlear stimulator (ics) device was caused by a loss of a hermetic seal.This is concluded from internal visual inspection results.Moisture admitted into this device through this leak resulted in the corrosion of the resistive film material and the shift in some resistance values.This ultimately caused the device to cease functioning.This older device configuration is no longer manufactured.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|