• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD (SUZHOU) BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD (SUZHOU) BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 383033
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/20/2022
Event Type  malfunction  
Manufacturer Narrative
Initial reporter phone #: (b)(6).Investigation summary: in response to the event reported by your facility a device history review was conducted for lot number 1350193.Our records show that this is the only instance of this issue occurring in this production batch.According to the sampling plan applied for product performance, this lot was accepted and released without defects being noted during the final assembly or visual inspections.A sample could not be obtained for evaluation and testing; in lieu of the affected device, functional testing was performed on retention samples for this lot, the results of these show that the tested units performed within product specifications.Unfortunately, without the ability to investigate the affected unit our quality engineers were unable to determine the root cause for this complaint.Bd will continue to monitor this issue and encourages you to submit your sample for review.
 
Event Description
It was reported by the customer that the bd intima-ii¿ closed iv catheter system was loose.The following information was provided by the initial reporter, translated from chinese to english: the indwelling needle prn loosens, causing fluid and blood to leak into the patient and sheets.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BD (SUZHOU)
no. 5 baiyu road
suzhou industrial park
suzhou
Manufacturer (Section G)
BD (SUZHOU)
no. 5 baiyu road
suzhou industrial park
suzhou
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15822799
MDR Text Key307609554
Report Number3014704491-2022-00582
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/18/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number383033
Device Lot Number1350193
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/26/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/16/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-