Model Number 5392 |
Device Problems
Display or Visual Feedback Problem (1184); Pacing Problem (1439)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/11/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
It was found on analysis that the external pulse generator (epg) ventricular pace pulse noise displayed an error and failed during the final test ran on the epg test system.An error was also observed on the epg display.The epg was checked and cleaned.The internal wire was checked and it was found not to be pinched or exposed.The main printed circuit board (pcb) was replaced.All found defective parts were replaced and all other identified issues were resolved.The device then passed all final functional tests.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned.¿ these words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
The external pulse generator (epg) which returned into service subsequently tested out of specification during manufacturer¿s analysis.There was no patient involvement.
|
|
Manufacturer Narrative
|
Product analysis update: further analysis of the external pulse generator (epg) was performed.The main board was assembled into a g olden unit.Upon functional test ran on automated test console the device failed ventricular pace output noise test measuring outside the requirements of ventricular pace pulse noise test specification.Upon benchtop analysis it was powered on and off five times and the reported error was not observed.Ventricular pace pulse noise was measured using o-scope and was found to be within the requirements of the ventricular pace pulse noise test specification.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|