Model Number 03.501.080 |
Device Problems
Mechanical Problem (1384); Device Slipped (1584)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during an unknown procedure on an unknown date in 2022, the sales consultant found that the external zipfix gun in question would not tighten.A back-up device was utilized, and the procedure was completed successfully.No further information was provided.This report is for an unk - zipfix implants.This is report 1 of 1 for (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Date of event: event occurred at an unknown date in 2022.This report is for an unknown zipfix implant/unknown lot.Part and lot numbers are unknown; udi number is unknown.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Initial reporter is a synthes employee.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Part number: 03.501.080.Lot number: 630p288.Manufacturing site: haegendorf.Release to warehouse date: 14.Mar.2022.Service and repair evaluation: the customer reported that 03.501.080, application instrument for snal zipfix would not tighten.The repair technician reported that the handle binds when fully engaged and the device failed to operate smoothly.Lube/oil/clean is the reason for repair.The cause of the issue is unknown.The item was repaired per the inspection sheet, passed synthes final inspection on 06-feb-2023 and will be returned to the customer upon completion of the service and repair process.The evaluation was confirmed.The device was deemed serviceable and will be returned to the customer, no design or manufacturing issues were identified therefore it has been determined that no corrective and/or preventative action is proposed.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
|
|