• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD SAF-T-INTIMA¿ IV CATHETER SAFETY SYSTEM; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD SAF-T-INTIMA¿ IV CATHETER SAFETY SYSTEM; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 383323
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/26/2022
Event Type  malfunction  
Event Description
It was reported that the bd saf-t-intima¿ iv catheter safety system prn was found loose after retracting the needle.The following information was provided by the initial reporter: "customer is reporting safety cap loose that controls blood flow when needle is retracted after catheter is place.".
 
Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Manufacturer Narrative
H6: investigation summary: a device history review was conducted for lot number 1261705.Our records show that this is the only instance of this issue occurring in this production batch.According to the sampling plan applied for product performance, this lot was accepted and released without defects being noted during the final assembly or visual inspections.A sample could not be obtained for evaluation and testing; in lieu of the affected device, functional testing was performed on retention samples for this lot, the results of these show that the tested units performed within product specifications.Unfortunately without the ability to investigate the affected unit our quality engineers were unable to determine the root cause for this complaint.H3 other text : see h10.
 
Event Description
It was reported that the bd saf-t-intima¿ iv catheter safety system prn was found loose after retracting the needle.The following information was provided by the initial reporter: "customer is reporting safety cap loose that controls blood flow when needle is retracted after catheter is place.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD SAF-T-INTIMA¿ IV CATHETER SAFETY SYSTEM
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX 
Manufacturer (Section G)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX  
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15829196
MDR Text Key304556181
Report Number9610847-2022-00437
Device Sequence Number1
Product Code FOZ
UDI-Device Identifier00382903833238
UDI-Public(01)00382903833238
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K923702
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/18/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number383323
Device Lot Number1261705
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/29/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/18/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-