Catalog Number 05942861041
Device Problems Incorrect Measurement (1383); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/31/2022
Event Type  malfunction  
Manufacturer Narrative
The customer's meter and test strips were requested for investigation.The product has not been received at this time.If the product is returned in the future, a follow-up report will be submitted.On a regular basis, accu-chek inform strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.
 
Event Description
The initial reporter stated they received discrepant glucose results for one patient tested with accu-chek inform ii meter serial number (b)(4).A sample from the patient was tested using the meter, resulting in a glucose value of 432 mg/dl.Within 15 minutes of initial testing, the patient was tested with the meter two additional times, each time resulting in a value of 117 mg/dl.
 
Manufacturer Narrative
The customer's meter and test strips were returned for investigation where they were tested using a retention battery and retention controls.Control ranges: level 1 = 30-60 mg/dl, level 2 = 261-353 mg/dl.Results: level 1: 44 mg/dl, 46 mg/dl, 44 mg/dl.Level 2: 305 mg/dl, 300 mg/dl, 303 mg/dl.All returned results are within the acceptable range.No information was provided that would point to a cause for the result discrepancy.Medwatch fields d9 and h3 have been updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCU-CHEK ® INFORM II TEST STRIPS
Type of Device
BLOOD GLUCOSE MONITORING TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15835519
MDR Text Key307185356
Report Number1823260-2022-03709
Device Sequence Number1
Product Code LFR
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
K121679
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/21/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2023
Device Catalogue Number05942861041
Device Lot Number670134
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/16/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-