• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROVENTION, INC. WEB SL; INTRASACCULAR DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION, INC. WEB SL; INTRASACCULAR DEVICES Back to Search Results
Model Number UNKNOWN
Device Problems Detachment of Device or Device Component (2907); Premature Separation (4045)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
A visual inspection of the device captured in this file could not be performed as a physical device was not returned for evaluation, nor were any images of the device provided in place of a device return.Procedure and medical imaging was not provided for this investigation.Without the return and physical evaluation of the device, the investigation cannot definitively determine if a condition existed that would have caused or contributed to the reported event.Without imaging, the investigation cannot verify the event occurred as described, nor could the investigation definitively determine the cause of the reported event.Based on a review of the device¿s risk documentation, the reported event did not indicate there were any potential or new manufacturing, design, quality, or other systemic issues, or non-conformances.The complaint code is monitored through the trending process; corrective action is determined, as needed, through this process.Investigations of historic complaint files with similar complaint category coding are recorded in the complaint handling system; without the ability to perform and analysis of the device, this investigation cannot identify with certainty any potential root causes.A search for non-conformances associated with this part/lot number combination could not be performed as this information was not available at the time this investigation was performed.Since the batch number for this product was not provided at the time of this investigation, a review of the complaint handling system could not be performed to determine if previous complaints have been submitted with a similar reported issue with this batch number.Potential complications include but are not limited to the following: hematoma at the site of entry, aneurysm rupture, emboli, vessel perforation, parent artery occlusion, hemorrhage, ischemia, vasospasm, clot formation, device migration or misplacement, premature or difficult device detachment, non-detachment, incomplete aneurysm filling, revascularization, post-embolization syndrome, and neurological deficits including stroke and death.Procedure: detachment of the device.The detachment control device is pre-loaded with batteries and will activate when the delivery device is properly connected.32.Verify that the rhv is firmly locked around the delivery device before attaching the detachment control device to ensure that the embolization device does not move during the connection process.33.Ensure that the delivery device gold connectors are clean and free from blood or contrast.If necessary, wipe the connectors with sterile water and dry before connecting.34.Insert the proximal end of the delivery device into the detachment control device.When the delivery device is properly connected, the light will flash green and an intermittent tone will be heard.35.Verify the embolization device position before pressing the detachment button.36.Push the detachment button.During firing, the light should be solid green and the beep should be continuous.37.Verify detachment by first loosening the rhv valve, then pulling back slowly on the delivery device and verifying that there is not embolization device movement.If the embolization device does not detach, push the detachment button again.If the device is still not detached, obtain a new detachment control device and attempt detachment up to two additional times.If it does not detach, remove the delivery device.38.Verify the position of the embolization device angiographically through the guide catheter.39.Prior to removing the microcatheter from the treatment site, place an appropriately sized guidewire completely through the microcatheter lumen to ensure that no part of the embolization device remains within the microcatheter.The physician has the discretion to modify the device deployment technique based on the complexity and variation in embolization procedures.Any modifications must be consistent with the previously described procedures, warnings, precautions and patient safety information in these instructions for use.The physical device was not available for evaluation to determine if a condition existed that would have caused or contributed to event.Supplemental imaging was also unavailable for review; without imaging, the investigation cannot verify the event occurred as described, nor could the investigation definitively determine the cause of the reported event.This information may be updated if additional information is provided at a later date.
 
Event Description
It was reported that the web did not release from the delivery pusher and unintentionally detached in an unintended location.No patient harm or intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WEB SL
Type of Device
INTRASACCULAR DEVICES
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise
aliso viejo, CA 92656
7142478000
MDR Report Key15838145
MDR Text Key307992123
Report Number2032493-2022-00531
Device Sequence Number1
Product Code OPR
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P170032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/21/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/04/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-