• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRICATH® FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRICATH® FOLEY CATHETER Back to Search Results
Catalog Number 0165L16
Device Problem Dent in Material (2526)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/03/2022
Event Type  malfunction  
Event Description
It was reported that foley catheter balloon was not symmetrical.It was also stated that a new biocath was reinserted.
 
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that foley catheter balloon was not symmetrical.It was also stated that a new biocath was reinserted.
 
Manufacturer Narrative
The reported event was inconclusive ¿ poor sample condition.The reported failure was able to be reproduced.It is unknown whether the device had met relevant specifications.The product was used for urological care.Based on the photos, it was observed that the catheter balloon was asymmetrically inflated.However, it is unknown if the balloon was under inflated which could also cause the asymmetrical balloon issue.Besides, no functional testing could be performed as only photo sample is provided.Therefore, the reported event is inconclusive due to poor sample condition.The potential root cause for this failure mode could be user related (example: under-inflation during use)/ assembly of the balloon to the shaft/ uneven thickness of balloon.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "warning: do not use ointments or lubricants having a petrolatum base.They will damage latex.Visually inspect the product for any imperfections or surface deterioration prior to use.¿ use luer tip syringe to inflate with stated ml of sterile water.Or ¿ for pre-filled products, remove clip and squeeze reservoir to inflate with stated ml of sterile water.For urological use only.To deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If necessary, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.This is a single use device.Do not resterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient." h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRICATH® FOLEY CATHETER
Type of Device
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key15843386
MDR Text Key306815793
Report Number1018233-2022-08880
Device Sequence Number1
Product Code EZC
Combination Product (y/n)N
Reporter Country CodeSN
PMA/PMN Number
K910846
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/22/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0165L16
Device Lot NumberMYFZ2329
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/03/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/24/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/18/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-