MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number HL 20
Device Problem Failure to Pump (1502)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/12/2022
Event Type  malfunction  
Manufacturer Narrative
It was reported that the hl20 device displayed the error message: "runaway." no harm to any person was reported.A getinge field service technician was onsite and investigated the unit in question.It was found that the tacho strobe foil on the otical tacho board was damaged and needs to be replaced.Thus the reported failure could be confirmed.Device was manufactured in 2008-04-25.The review of the non-conformities during the period of 2008-04-25 to 2022-11-23 does not show any non-conformity in regard to the reported product and failure.There is no indication that manufacturing issues occurred during this time, thus production related influences are unlikely.An investigation of the tacho strobe foil is not possible as it is always damaged during removing of it.The most probable root cause could be determined as a mechanical damage of the marks on the tacho strobe foil.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
It was reported that the hl20 device displayed the error message: "runaway." no harm to any person was reported.Complaint id # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART LUNG MACHINE
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key15850366
MDR Text Key304212659
Report Number8010762-2022-00464
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K943803
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Consumer,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/23/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL 20
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/12/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/25/2008
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-