MAUDE Adverse Event Report: TOSOH HI-TEC, INC. AIA-900; FLUOROMETER, FOR CLINICAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH HI-TEC, INC. AIA-900; FLUOROMETER, FOR CLINICAL USE Back to Search Results
Model Number AIA-900
Device Problem Priming Problem (4040)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/20/2022
Event Type  malfunction  
Manufacturer Narrative
A field service engineer (fse) was at the customer¿s site to address the reported event.Fse confirmed the complaint by reviewing the error logs and was unable to reproduce the error because it was intermittent.While troubleshooting, fse visually saw a sign of leakage on top of the incubator cover and replaced the wash syringe.Fse proactively replaced both wash probes that was located around the metal suction tube due to visually seeing leakage.Fse repaired and validated the analyzer.No further action required by field service.The aia-900 analyzer is functioning as expected.A 13-month complaint history review and service history review for similar complaints was performed for the serial number: (b)(4) from 27sep2021 through aware date 27oct2022.There were four similar complaints identified during the search period including this case.The aia-900 operator's manual under section12: error messages states the following: (2239) bf probe 1 purge failure.Cause: the overflow sensor 1 s132 failed to detect liquid after the washer was discharged.Action: it is conceivable that air has entered the washer tube.Prime the tube and bleed off the air.Check the remaining quality of washer, check to see if there is air in the washer tube, and check s132, the discharge solenoid valve sv150, and the washer tube.The aia-900 operator's manual under section12: error messages states the following: (2240) bf probe 2 purge failure.Cause: the overflow sensor 2 s133 failed to detect liquid after the washer was discharged.Action: it is conceivable that air has entered the washer tube.Prime the tube and bleed off the air.Check the remaining quality of washer, check to see if there is air in the washer tube, and check s133, the discharge solenoid valve sv151, and the washer tube.The most probable cause of the reported event was due to the faulty wash syringe.
 
Event Description
A customer reported error messages ¿2239 bf probe 1 purge failure and 2240 bf probe 2 purge failure¿ on the aia-900 analyzer.The analyzer is down.A field service engineer (fse) was dispatched to address the reported event which resulted in a delayed reporting of patient samples for creatine kinase mb isoenzyme (ck-mb) and cardiac troponin i (ctnl 2).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Manufacturer Narrative
Correction section d10.The wash syringe and wash probes were not received by the instrument service center because they were disposed of onsite by a field service engineer due to possible contamination.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-900
Type of Device
FLUOROMETER, FOR CLINICAL USE
Manufacturer (Section D)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA  7460042
Manufacturer (Section G)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA   7460042
Manufacturer Contact
jorge porras
shiba-koen first building
3-8-2 shiba
minato-ku 10586-23
JA   1058623
6506368314
MDR Report Key15854477
MDR Text Key307798124
Report Number3004529019-2022-00190
Device Sequence Number1
Product Code KHO
UDI-Device Identifier04560189283992
UDI-Public04560189283992
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/23/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-900
Device Catalogue Number022930
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/04/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/07/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-