Model Number UNKNOWN
Device Problem Structural Problem (2506)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/28/2022
Event Type  malfunction  
Manufacturer Narrative
Initial reporter phone #:(b)(6).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that while using the bd kiestra¿ inoqula+¿ tla that the barcoda hood was falling.The following information was provided by the initial reporter: replace barcoda hood gas springs.
 
Manufacturer Narrative
H.6 investigation summary: bd technical services reported a worn-out gas spring on instrument bd kiestra inoqula tla (material:447213, serial number: (b)(6)).Springs were replaced with new set which has confirmed the complaint.Root cause is defined as faulty springs, and replacement with new set is initiated as a corrective action.Design history record (dhr) review is not required for this complaint, the complaint was evaluated via other elements of the investigation.The results of this evaluation have not identified any new hazards, new risks, or specific trends.
 
Event Description
It was reported that while using the bd kiestra¿ inoqula+¿ tla that the barcoda hood was falling.The following information was provided by the initial reporter: replace barcoda hood gas springs.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD KIESTRA¿ INOQULA+¿ TLA
Type of Device
MICROBIAL SPECIMEN INOCULATION/STREAKING INSTRUMENT
Manufacturer (Section D)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15859668
MDR Text Key307089989
Report Number1119779-2022-01423
Device Sequence Number1
Product Code JTC
UDI-Device Identifier00382904472139
UDI-Public00382904472139
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/23/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberUNKNOWN
Device Catalogue Number447213
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/01/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/06/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-