• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON MONARCH IOL DELIVERY SYSTEM, CARTRIDGE; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON MONARCH IOL DELIVERY SYSTEM, CARTRIDGE; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL) Back to Search Results
Model Number ASKU
Device Problems Device Damaged by Another Device (2915); Appropriate Term/Code Not Available (3191)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history record could not be reviewed because the reporting facility did not provide a valid lot number or any identification traceable to the manufacturing documentation.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Event Description
An account manager reported that several lenses were dinged.Additional information has been requested and received stated that cartridge has two straight lenses that loaded improperly and the plunger slipped past the edge and end up hitting the central optic and the optic needing to be explanted.Additional information has been requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONARCH IOL DELIVERY SYSTEM, CARTRIDGE
Type of Device
FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15861675
MDR Text Key307723098
Report Number1119421-2022-02472
Device Sequence Number1
Product Code MSS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P063155
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 11/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/24/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberASKU
Device Catalogue NumberASKU
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/27/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CLAREON ASPHERIC UV ABSORBING IOL; PROVISC OPHTHALMIC VISCOSURGICAL DEVICE
-
-