Catalog Number 101025
Device Problem Disconnection (1171)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported within minutes of starting treatment with an hd cartridge line, the dialysis machine alarmed an unspecified alarm.It was observed that the arterial blood line was disconnected from the arterial injection port.No blood loss was reported.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The actual device was not received; however, one (1) sample picture was provided for evaluation confirming a disconnection between arterial patient line and access site red.The reported condition was verified.The cause of the condition was determined to be a manufacturing issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HD CARTRIDGE LINES
Type of Device
SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - TIJUANA BAJA
blvd pacifico 10014
parque industrial pacifico
tijuana baja california CP 22 643
MX   CP 22643
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key15875256
MDR Text Key304515931
Report Number8030638-2022-00032
Device Sequence Number1
Product Code FJK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K070414
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 11/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/28/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number101025
Device Lot NumberMA012208524
Was Device Available for Evaluation? No
Date Manufacturer Received11/23/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/02/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-