W. L. GORE & ASSOCIATES, INC. GORE® ACUSEAL VASCULAR GRAFT; PROSTHESIS, VASCULAR GRAFT Back to Search Results
Device Problems Peeled/Delaminated (1454); Material Puncture/Hole (1504)
Patient Problems Hematoma (1884); Stenosis (2263)
Event Date 11/09/2022
Event Type  Injury  
Event Description
The following publication was reviewed: title: experience of acuseal and troubleshooting for mid-to-long term follow-up.Source: journal of japanese society for dialysis therapy 2022: 55(suppl.1) p.618.This report includes 6 cases of acuseal implantation in one clinic since 2017.The primary patency rate was 83.3% at 1 year and 50% at 3 years, and the secondary patency rate was 100% at both 1 and 3 years.Hematoma formation associated with wall penetration due to early puncture was observed during the course of follow-up so that the ultrasound-guided puncture was performed.As mid-to-long term complications, three patients had vascular stenosis due to loss of wall material from intimal structures of grafts.The material was removed using biopsy forceps through a sheath.The damage to the graft wall was considered to be occurred by concentration of puncture sites.
 
Manufacturer Narrative
Cbas® heparin surface incorporates carmeda heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.Title: experience of acuseal and troubleshooting for mid-to-long term follow-up.Source: journal of japanese society for dialysis therapy 2022: 55(suppl.1) p.618 physician: kyuichi furuhashi.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Manufacturer Narrative
Corrected medical device problem code.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® ACUSEAL VASCULAR GRAFT
Type of Device
PROSTHESIS, VASCULAR GRAFT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL WEST B/P
1505 n. fourth street
flagstaff AZ 86004
Manufacturer Contact
jorja nackard
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key15878606
MDR Text Key304447635
Report Number2017233-2022-03516
Device Sequence Number1
Product Code DSY
Combination Product (y/n)Y
Reporter Country CodeJA
PMA/PMN Number
K130215
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Literature
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 02/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/29/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/09/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Other;
-
-